Besides the standard stainless steel version, the O2X1 is also available in two other packages:
* The first package is a lower-cost package using Delrin for the wetted parts of the transmitter. This package has the same 1/8 in (3.175 mm) NPTF processes connections as the standard stainless steel O2X1 and is suitable for all ranges except the lowest PPM oxygen range.* The second package makes the Delrin design suitable for ambient air oxygen monitoring by providing a diffusion cut-out in the lower manifold to expose the sensor. This package comeswith an aluminum mounting bracket for mounting the O2X1 to any convenient wall location.
- Measures oxygen from 1.0 ppm to 25 percent
- Six field-programmable ranges
- Repeatability ± 1 percent of span
- Self-contained cell requires no maintenance
- 4 to 20 mA, loop powered, 9 to 28 VDC

- Measures oxygen from 0.01 to 100 percent in gases.
- No moving parts; maintenance-free
- Corrosion-resistant design
- Automatic background-gas compensation
- Use as stand-alone transmitter or in conjuction with TMO2D or XDP display packages

- Measures oxygen from 0.1 to 25 percent in gases
- No moving parts; maintenance-free
- Corrosion-resistant design
- Automatic background-gas compensation
- Automatic process pressure compensation

- Optimized reaction processes
- Improved effi ciency and increased process up-time
- Greater safety assurance
- Reduced total cost of ownership
- Reduced installation cost for hazardous area locations
- Remote monitoring

The rugged XDP display/controller is microprocessor-based and certified to Class 1, Division1, Groups B,C,D and ATEX EEx d IIC T6. Most importantly, the unit features an infrared, thru-glass touch keypad, which allows complete access to all of the XDP user programmable features including output type, range selection, automatic calibration, and error handling. The IR keypad feature allows field engineers to interact with the analyzer through the front display without the need of opening the enclosure and thus avoiding the need for traditional hot work permits.
- Explosion-proof display/controller
- Compatible with XMO2 oxygen transmitter, XMTC thermal conductivity gas transmitter or any 4-20mA output transmitter
- Universal input power supply
- IR through-the-glass keypad
- Fault alarm
- Auto-calibration alarm relays available for general applications
- Optional software specifically designed for use with the XMTC for hydrogen analysis on hydrogen cooled generator applications. Measure all three gas cycles for hydrogen cooled generators:
- 0 to 100 percent hydrogen in air
- 0 to 100 percent hydrogen in carbon dioxide
- 0 to 100 percent air in carbon dioxide